Atlas FDA

Magseed Pro 7 cm soft tissue marker. Model Number: MS2-17-1-07. Product Description: The Endomag Magseed Pro Magnetic Ma

FDA device recall Z-0316-2026 · posted 2025-10-24 · Open, Classified

Recall details

Recalling firm
ENDOMAGNETICS LTD
Reason for recall
Potential for contamination with cotton fibers.
Action taken
Endomagnetics, Ltd notified consignees on about 10/01/2025 via certified mail and email through 3rd party IQVIA. Consignees were instructed to examine their inventory and cease use of those units, quarantine any affected units, complete and return the response form, forward the notification to all affected personnel and customers if further distributed, and post a copy of the notice in a visible area for awareness and keep a copy for your records.
Root cause
Process control
Status
Open, Classified
Product code
NEU
Quantity affected
1,760 units
Distribution
Worldwide - US Nationwide distribution in the states of CA, CO, CT, DC, FL, GA, IA, ID, IN, KY, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NY, OH, TX, WA and the country of Canada.
Date initiated
2025-10-01
Date posted
2025-10-24
Location
Cambridge, United Kingdom

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
Magseed Pro Magnetic Marker System (K232865)Endomagnetics Ltd.,2024-06-28