Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System. Maquet GmbH &
FDA device recall Z-2097-2015 · posted 2015-07-17 · Terminated
Recall details
- Recalling firm
- Maquet Cardiovascular Us Sales, Llc
- Reason for recall
- There is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in rare cases, MAGNUS column provides wrong position data for the longitudinal movement leading to a collision.
- Action taken
- Maquet sent a field correction letter dated June 8, 2015, via Fed Ex. Customers are advised of the issue with the device and provided instructions. Questions or additional information can be obtained from the local Maquet representative or Maquet Technical Support at 1-888-627-8383 (option 3 followed by option 1) Monday through Friday between the hours of 8:00 am and 6:00 pm EST.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- FQO
- Quantity affected
- 12 units
- Distribution
- Worldwide Distribution - US Nationwide Distribution and to the countries of : France, Australia, Belgium, Brazil, Switzerland, Chile, China, Germany, Spain, United Kingdom, Hong Kong, Ireland, Israel,
- Date initiated
- 2015-06-10
- Date posted
- 2015-07-17
- Date terminated
- 2016-04-18
- Location
- Wayne, NJ