Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System. Maquet GmbH &

FDA device recall Z-2097-2015 · posted 2015-07-17 · Terminated

Recall details

Recalling firm
Maquet Cardiovascular Us Sales, Llc
Reason for recall
There is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in rare cases, MAGNUS column provides wrong position data for the longitudinal movement leading to a collision.
Action taken
Maquet sent a field correction letter dated June 8, 2015, via Fed Ex. Customers are advised of the issue with the device and provided instructions. Questions or additional information can be obtained from the local Maquet representative or Maquet Technical Support at 1-888-627-8383 (option 3 followed by option 1) Monday through Friday between the hours of 8:00 am and 6:00 pm EST.
Root cause
Device Design
Status
Terminated
Product code
FQO
Quantity affected
12 units
Distribution
Worldwide Distribution - US Nationwide Distribution and to the countries of : France, Australia, Belgium, Brazil, Switzerland, Chile, China, Germany, Spain, United Kingdom, Hong Kong, Ireland, Israel,
Date initiated
2015-06-10
Date posted
2015-07-17
Date terminated
2016-04-18
Location
Wayne, NJ