Magnetom Espree with Swiveling OR (Operating Room) Table, Magnetic Resonance Imaging System,Siemens Medical Solutions US

FDA device recall Z-0518-2008 · posted 2008-02-05 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc
Reason for recall
Table may experience a deadlock situation
Action taken
The recalling firm issued a Customer Safety Advisory to affected customers per Update Instructions MR031/07/S. The letter informs customers of the potential issue and provides instructions to avoid its occurrence. The recalling firm has also released the software update to correct the issue via Update Instructions MR034/07/S. Notifications were issued on or about November 5, 2007.
Root cause
Software design
Status
Terminated
Product code
LNH
Quantity affected
4 units
Distribution
The products were shipped to medical facilities in TX and NY.
Date initiated
2007-11-05
Date posted
2008-02-05
Date terminated
2008-02-05
Location
Malvern, PA