M3536A HeartStart MRx with Wireless Link Upgrade (861289) The HeartStart MRx is for use for the termination of ventricul

FDA device recall Z-1108-2014 · posted 2014-02-26 · Terminated

Recall details

Recalling firm
Philips Medical Systems, Inc.
Reason for recall
Wireless Link may be unable to transmit data when configured for a specific data flow.
Action taken
Philips Healthcare sent an Urgent Medical Device Correction letter to all consignees on February 18, 2014. The Letter informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. Philips is initiating a correction to affected devices. Philips will release a new version of Wireless Link software V2.0.0.3 sub1 that corrects the issue. The correction will be provided free of charge to all units affected by this issue. A Philips Healthcare representative will contact customers with affected devices to arrange for installation of the software upgrade. Customers with questons were instructed to call 1-800-722-9377. For questions regarding this recall call 800-722-9377.
Root cause
Software design
Status
Terminated
Product code
MKJ
Quantity affected
220 devices
Distribution
Worldwide Distribution - USA including AZ, CA, FL, IL, IN, KS, KY, MA, MD, MI, MN, MO, NC, NY, OH, OK, OR, PA, SC, TN, TX,and VA and Internationally to Australia, Norway, and Spain.
Date initiated
2014-02-18
Date posted
2014-02-26
Date terminated
2015-11-12
Location
Andover, MA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
HEARTSTART MRX WITH Q-CPR OPTION, MODELS M3535A OR M3536A (K051134)Philips Medical Systems2005-09-08