LX3 Floor Stand for Endure Luxite Envision Optical System; Endure LuxOR Q-Vue. Product Usage: The LX3 Microscope Floor S

FDA device recall Z-2448-2015 · posted 2015-08-21 · Terminated

Recall details

Recalling firm
Alcon Research, Ltd.
Reason for recall
Alcon is initiating a voluntary recall of all LX3 Microscope Floor Stands after receiving reports concerning loss of functionality due to issues with the cabling between the power, host and illumination modules, potentially causing the system to display a system message, and/or resulting in illumination failure.
Action taken
Alcon sent an Urgent Medical Device Correction letter dated August 11, 2015 to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact Alcon at 800-862-5266.
Root cause
Device Design
Status
Terminated
Product code
HRM
Quantity affected
688 units
Distribution
Worldwide Distribution - US Nationwide and the countries Argentina, Australia, Austria, Belarus, Belgium, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Dominican Repub
Date initiated
2015-08-11
Date posted
2015-08-21
Date terminated
2016-11-04
Location
Fort Worth, TX