LX3 Floor Stand for Endure Luxite Envision Optical System; Endure LuxOR Q-Vue. Product Usage: The LX3 Microscope Floor S
FDA device recall Z-2448-2015 · posted 2015-08-21 · Terminated
Recall details
- Recalling firm
- Alcon Research, Ltd.
- Reason for recall
- Alcon is initiating a voluntary recall of all LX3 Microscope Floor Stands after receiving reports concerning loss of functionality due to issues with the cabling between the power, host and illumination modules, potentially causing the system to display a system message, and/or resulting in illumination failure.
- Action taken
- Alcon sent an Urgent Medical Device Correction letter dated August 11, 2015 to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact Alcon at 800-862-5266.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- HRM
- Quantity affected
- 688 units
- Distribution
- Worldwide Distribution - US Nationwide and the countries Argentina, Australia, Austria, Belarus, Belgium, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Dominican Repub
- Date initiated
- 2015-08-11
- Date posted
- 2015-08-21
- Date terminated
- 2016-11-04
- Location
- Fort Worth, TX