LUMEX A Graham-Field Brand, Bed End Sert. Low/Full-electric, Plastic Panels, Item 690-7604-000 Manufactured for : GF Hea
FDA device recall Z-1536-2012 · posted 2012-05-11 · Terminated
Recall details
- Recalling firm
- GF Health Products, Inc./d.b.a.Basic American Metal Products
- Reason for recall
- GF Health Products, Inc., is conducting a recall on the Lumex Patriot Full Electric Homecare "Bed Ends" manufactured 3/27/09, 5/9/09, 5/30/09 and 6/20/09. The supplier deviated from the print specifications of a machined pin to a sheared pin. This deviation has the potential to provide room for the pin to move during use. If failure were to occur, the bed could descend rapidly causing non-typic
- Action taken
- GF Health Products, Inc. sent an "URGENT CORRECTION AND REMOVAL" letter dated April 17, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to immediately cease use of the affected product, secure the product, and to prevent the product from further distribution to customers and end users. Contact the firm at 1-800-556-6508 for questions regarding this notice.
- Root cause
- Component change control
- Status
- Terminated
- Product code
- LLI
- Quantity affected
- 3095 pairs of bed ends
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of AL, AZ, AR, CA, CO, CT, FL, GA, HI, ID, IL IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, S
- Date initiated
- 2012-04-17
- Date posted
- 2012-05-11
- Date terminated
- 2013-07-15
- Location
- Fond Du Lac, WI