Low Profile U-Joint Driver for the SynFix Mini-Open System, Catalog number 03.802.231. The Synthes SynFix LR is indicate
FDA device recall Z-0931-2009 · posted 2009-01-30 · Terminated
Recall details
- Recalling firm
- Synthes Spine
- Reason for recall
- screwdriver tip may break off and lodge in the screw head
- Action taken
- The recalling firm emailed all Spine Sales Consultants on 11/20/08. They were instructed to immediately examine their stock and discuss appropriate technique with surgeon users. In addition, a Medical Device Correction letter dated 11/19/08 was issued via USPS certified mail to the medical facilities. This initial letter stated the issue and requested the users take the following actions: discard all SynFix Technique Guides and replace with the updated literature contained in the mailing packet with the letter, and is identified by code J7022-E (Rev. E); to ensure receipt and understanding of the information, the Verification Section of the letter needs to be completed and returned to Synthes by either fax: 610-719-5120 or scan/e-mail: lewis.lynne@synthes.com. An Urgent: Medical Device Recall letter, dated 1/16/2009, was also sent out. This latter letter requested that users examine their stock immediately and remove any of the affected devices from their inventory. Customers are also to call Synthes at 1-800-479-6329 to obtain a Return Authorization Number; complete the Verification Section; and return the letter with the recalled product to the address listed. There are also actions for those who do not have the affected product. Questions should be directed to 800-620-7025 x 5453 or 610-719-5453.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- MAX
- Quantity affected
- 381 units
- Distribution
- Class II Recall - Worldwide Distribution --- USA including states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, PA ,SC, TN,
- Date initiated
- 2008-11-20
- Date posted
- 2009-01-30
- Date terminated
- 2009-10-27
- Location
- West Chester, PA