LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrogr
FDA device recall Z-0229-2026 · posted 2025-10-21 · Open, Classified
Recall details
- Recalling firm
- Synthes (USA) Products LLC
- Reason for recall
- Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.
- Action taken
- DePuy Synthes notified consignees via email and mailed letter on 09/26/2025. Consignees were instructed to examine inventory and quarantine affected units, arrange for the return of any affected units, complete and return the Business Response Form, and place the notification in a visible area for awareness. If units were further distributed within your facility or to customers, please notify those customers and personnel.
- Root cause
- Process control
- Status
- Open, Classified
- Product code
- JDS
- Quantity affected
- 122 units
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, MS, NJ, NM, NY, OH, TN, TX and the country of Canada.
- Date initiated
- 2025-09-26
- Date posted
- 2025-10-21
- Location
- West Chester, PA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| DePuy Synthes Retrograde Femoral Nail Advanced System, DePuy Synthes Locking Screws for Medullary Nails, 5.0 mm (K201346) | Synthes (USA) Products, LLC | 2020-09-23 |
| DePuy Synthes Tibial Nail Advanced, Depuy Synthes Locking Screws Medullary Nails, 4.0 and 5.0mm (K201336) | Synthes (USA) Products, LLC | 2020-09-03 |