LOGIX Order Entry (OE) Software, all versions, catalog 2M8410; Baxter Healthcare, Deerfield, IL 60015; The software is i

FDA device recall Z-2291-2011 · posted 2011-05-25 · Terminated

Recall details

Recalling firm
Baxter Healthcare Corp.
Reason for recall
The LOGIX OE automated backup does not retain the three most recent backup files and instead will only retain the last three backups of the previous year. The software stores backup files in numeric order rather than chronological date order.
Action taken
Baxter Healthcare Corp. sent an URGENT DEVICE CORRECTION letter dated March 29, 2011, via first class mail to all affected customers. The letter identified the product, the problem and the action to be taken by the customer. It was discovered there is an issue with the LOGIX OE automated backup not retaining the three most recent backup files and instead will only retain the last three backups of the previous year. The accounts were requested to turn off the automated backup function on the LOGIX OE Software, and perform manual backups following the instructions included in the Urgent Device Correction letter, and retain a copy of the letter at each computer on which LOGIX OE Server is installed. The customer were also requested to complete the attached customer reply form and fax it to Baxter at the number provided on the form. Any questions regarding the communication were directed to Baxter's Medical Information Services at 1-800-422-2751.
Root cause
Software design
Status
Terminated
Product code
LHI
Quantity affected
84 units
Distribution
Worldwide Distribution -- USA including Puerto Rico, and the country Canada.
Date initiated
2011-03-29
Date posted
2011-05-25
Date terminated
2017-01-05
Location
Round Lake, IL