Liquid Mouse Monoclonal Antibody CD15 (NCL-L-CD15), a liquid tissue culture supernatant containing 15 mM sodium azide as
FDA device recall Z-1710-2014 · posted 2014-06-04 · Terminated
Recall details
- Recalling firm
- Leica Microsystems, Inc.
- Reason for recall
- Product does not function as intended up to the expiry date on the product labeling. There is a link between the age of the product and staining intensity.
- Action taken
- Leica sent an Urgent Field Safety Correction Notice letter dated January 10, 2014 was sent ot all affected customers. The notice included instructions for recipients to: 1) destroy any unused or partially used NCL-L-CD15 (lots 6015073 and 6017936) reagent or confirm that the products are no longer in stock; 2) indicate on the attached Field Safety Correction Notice Acknowledgement Fom which action was taken; and, 3) contact the local Leica represetnative immediately if a replacement is required. Customers with any concerns about this notice can contact Leica via e-mail at LMGRA@leica-microsystems.com.
- Root cause
- Incorrect or no expiration date
- Status
- Terminated
- Product code
- NJT
- Quantity affected
- 145 units
- Distribution
- Worldwide Distribution - US Nationwide in the states of CA, CT FL, IL, MO, MS, NC, NY, PA, and the countries of Argentina, Australia, Brazil, Colombia, Italy, Malaysia, New Zealand, Nigeria, Poland, R
- Date initiated
- 2014-01-10
- Date posted
- 2014-06-04
- Date terminated
- 2017-09-05
- Location
- Buffalo Grove, IL