Lipase Reagent, 7D80. Used to measure lipase in human serum on the Architect system.
FDA device recall Z-0131-2015 · posted 2014-10-28 · Terminated
Recall details
- Recalling firm
- Abbott Laboratories, Inc
- Reason for recall
- Package inserts have incorrect SmartWash parameters for Triglyceride.
- Action taken
- Firm issued a Product Correction Letter to all Lipase reagent customers who received lots 43827UQ12, 43819UQ03, 44449UQ05, 44127UQ06, 44334UQ07, 43994UQ07, 44596UQ08, 44775UQ09, and 44930UQ11. The letter identified the affected product and provided an explanation of the problem, as well as, the patient impact and actions to be taken. The Product Correction Letter provided guidance for the laboratory to correctly configure the Architect cSystem's Lipase assay parameters. Customers are to retain a copy of the letter for their lab records, and forward a copy to any laboratories that the product was further distributed to. Questions should be directed to Customer Service at 1-877-4ABBOTT or your local customer service for customers outside of the US.
- Root cause
- Labeling Change Control
- Status
- Terminated
- Product code
- CHI
- Quantity affected
- 11916 units
- Distribution
- Worldwide Distribution: US (nationwide) and countries of: Algeria, Angola, Argentina, Armenia, Australia, Austria, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China,
- Date initiated
- 2014-04-07
- Date posted
- 2014-10-28
- Date terminated
- 2015-04-01
- Location
- Irving, TX