Linet Remote Handset for model Multicare, Eleganza 3 and Eleganza Smart, hospital bed. Product Usage: Positional hospita
FDA device recall Z-0738-2012 · posted 2012-01-13 · Terminated
Recall details
- Recalling firm
- Linet Americas Inc
- Reason for recall
- The premature failure of the Remote Handset buttons may result in unintended movement of the bed.
- Action taken
- Linet sent a notification letter dated December 9, 2011 to all affected customers. The letter identified the product, problem, and actions to be taken. The letter states that firm Linet has decided to do a voluntary field correction to replace the recalled product. The replacement will be completed by the territory Representative (Linet Employee) who will replace the affected product for each customer personally. The replacement takes approximately 30 seconds and does not require the patient to be moved. For questions or concerns call 704-248-5650.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- FNL
- Quantity affected
- 29,252 units
- Distribution
- Worldwide Distribution - USA (nationwide) including the states of: AL, FL, KY, LA, MA, NC, NY, OH and the country of: Canada
- Date initiated
- 2011-12-08
- Date posted
- 2012-01-13
- Date terminated
- 2012-07-20
- Location
- Charlotte, NC