Linear Medical Accelerator ARTISTE MV systems, when equipped with a syngo RT Therapist, Therapist Express Basic, syngo R
FDA device recall Z-1064-2011 · posted 2011-02-08 · Terminated
Recall details
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Reason for recall
- Potential for Mistreatment-- when treatment plans intended for non-ARTISTE machinese are loaded, if the plans contain unspecified beam parameters, attributes or unknown block codes, the therapy system may remove those parameters.
- Action taken
- The firm,Siemens, sent two Customer Advisory letters (TH004/09/S and TH008/09/S) contained in the Update Instructions to all consignees/customers with active devices. The letters describe the product, problem and action to be taken by the customers. The customers were instructed to follow the updated instructions. Siemens service representative will sign a document noted as CORRECTED upon completion of the software installation. If you have any questions, call 1-925-602-8157.
- Root cause
- Software design
- Status
- Terminated
- Product code
- IYE
- Quantity affected
- 151 units
- Distribution
- Worldwide distribution: USA and countries including: Germany, France, Poland, Norway, Australia, Malaysia, Spain, New Zealand, the UK, India, Republic of Korea, Canada, Ireland, Italy, Egypt, Belgium,
- Date initiated
- 2009-07-30
- Date posted
- 2011-02-08
- Date terminated
- 2011-04-12
- Location
- Concord, CA