Atlas FDA

Liko Sabina II EE Lift, Liko, Sweeden. The Sabina and Capella sit-to-stand lifts are especially designed for people who

FDA device recall Z-1692-2011 · posted 2011-03-17 · Terminated

Recall details

Recalling firm
Hill-Rom, Inc.
Reason for recall
The firm has received reports of injuries related to the Sabina Sit-in-Stand lifts. Of six injury reports, two were injuries in patints during use and four injuries were in non-patients when they tripped over the lift device and fell on the unit.
Action taken
The firm, Liko, issued two "URGENT FIELD SAFETY NOTICES" one dated January 27, 2011 and a follow-up notice dated February 9, 2011 to its consignees/customers. The notices described the product, problem and actions to be taken. The customers were instructed to keep the device in a separate unoccupied area when not in use; use extra care and support when lifting patients with the device; not allow unauthorized persons, especially children, to play around or operate the lift; examine the sling hooks daily to make sure they are secure, and to forward a copy of this notice to any other facility personnel they deem appropriate. The notices state the firm is investigating a design change to mitigate the risk of injuries with the device. If you have any questions concerning this request or the procedure as outlined, please contact Hill-Rom Technical Support at 800-445-3720.
Root cause
Device Design
Status
Terminated
Product code
FSA
Distribution
Worldwide distribution: USA including: AL, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK ,OR, PA, RI, SC, SD, TN ,TX, UT, VA, W
Date initiated
2011-01-27
Date posted
2011-03-17
Date terminated
2014-09-19
Location
Batesville, IN

Monitor recalls programmatically