Lifeshield, latex-free, Primary I.V. Pump Set with distal microbore patient line and gravity flow prevention valve, conv

FDA device recall Z-1352-2010 · posted 2010-04-15 · Terminated

Recall details

Recalling firm
Hospira Inc.
Reason for recall
Reports of difficulty in adjusting the flow rate, which could result in flow rate variances.
Action taken
Hospira sent the Urgent Device Recall letters dated 11/25/09 to their direct accounts of the affected products and requested these consignees to notify their customers of the recall. The accounts were informed that Hospira ceased manufacturing the I.V. gravity and pump administration sets with the CAIR clamp manufactured by Borla S.P.A. of Italy in March 2008, due to customer reports related to a difference in the zone of control used to adjust the flow compared to the original CAIR clamp design. To avoid the possibility of flow rate variances, Hospira is recalling all remaining sets with the Borla designed CAIR clamp. The accounts were requested to quarantine all inventory of the affected lots, complete the enclosed reply form indicating the amount of product being returned, fax a copy of the completed form to Stericycle at 1-877-377-5114, and include a copy with the returned goods to Stericycle at 2670 Executive Drive, Suite A, Indianapolis, IN 46241.
Root cause
Component design/selection
Status
Terminated
Product code
FPA
Quantity affected
14,833 sets
Distribution
Worldwide Distribution -- U.S., Puerto Rico, Dominican Republic, Bahamas, Virgin Islands, Argentina, Australia, Bahrain, Belgium, Canada, China, Denmark, Ecuador, EI Salvador, Finland, France, Germany
Date initiated
2009-11-24
Date posted
2010-04-15
Date terminated
2012-01-29
Location
Lake Forest, IL