LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software SUite v2.1 and v2.2
FDA device recall Z-1763-2025 · posted 2025-05-14 · Open, Classified
Recall details
- Recalling firm
- ICU Medical, Inc.
- Reason for recall
- Firm has identified Drug Library Management defects in the software: 1) DLM software does not allow user to create percentage dose rate or rate change values with certain limits, potentially resulting in over-delivery to patient. 2) An unauthorized user may modify and approve a drug library, potentially leading to incorrect program parameters being used for therapy.
- Action taken
- Firm began notifying consignees on May 2, 2025 via hand delivery of letters titled "Urgent Medical Device Correction." Issue 1: User receives an error if they attempt to define a Percentage Dose Change Increase limit between 60% and 99%. Customers were instructed to ensure all users or potential users of the software are made aware of the notification and proposed mitigations: Use Percentage Dose Change Increase values between 20% and 59% or between 100% and 500%. Pharmacists will not be able to use values from 60% to 99%. Issue 2: It is possible for an unauthorized user to bypass the reauthentication steps required to approve a drug library or delete a medication ruleset. Customers were instructed to ensure all users or potential users are made aware of the notification and proposed mitigation: Communicate with all users the need to log off after finishing a session before leaving the workstation and following cybersecurity best practices. ICU Medical is developing a LifeShield software update to correct both issues.
- Root cause
- Software Design Change
- Status
- Open, Classified
- Product code
- PHC
- Quantity affected
- 12 US, 3 OUS
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, FL, IA, KY, LA, MA, MS, NY, PA, TN, and VA. The country of Philippines.
- Date initiated
- 2025-05-02
- Date posted
- 2025-05-14
- Location
- Lake Forest, IL