Atlas FDA

LifeScan OneTouch Data Management Software Version 1.0, Part Number: 021-722-01(for OneTouch Select and UltraMini (Downl

FDA device recall Z-1317-2008 · posted 2008-04-08 · Terminated

Recall details

Recalling firm
Lifescan Inc
Reason for recall
Meter Temporarily Freezes-- A software compatibility issue may cause the blood glucose meter to cease operations and freeze temporarily.
Action taken
The firm sent an Urgent Recall letter to Health Care Professionals on 12/18/2007. The letter advised user to uninstall software drivers and provided options on how to uninstall. For further assistance, please contact LifeScan Customer Service at 1-866-679-9250.
Root cause
Software design
Status
Terminated
Product code
NBW
Quantity affected
260 units
Distribution
Nationwide Distribution including the states of AL, CA, IN, IO, MA, MD, MI, MS, NE, NY, NC, OR, PA, TN, TX, UT, VT, VA, WV, and WI.
Date initiated
2007-12-18
Date posted
2008-04-08
Date terminated
2008-07-18
Location
Milpitas, CA