LIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo RD rainbow Adt 8 SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-00
FDA device recall Z-1184-2025 · posted 2025-02-19 · Open, Classified
Recall details
- Recalling firm
- Physio-Control, Inc.
- Reason for recall
- Due to an error message that prevents users from utilizing carboxyhemoglobin saturation and methemoglobin saturation sensors on their monitor/defibrillator system.
- Action taken
- On 1/21/2025, the firm send an "URGENT: MEDICAL DEVICE RECALL" Letter via UPS 2-Day mail to customers to inform them that the LIFEPAK 15 V4 displays an error where "SpO2: Sensor does not support SpCO or SpMET". The error results in preventing users from utilizing LIFEPAK 15 V4 SpCO and SpMet functionalities when connected to specific lots of Masimo Rainbow Sensors. Customers are instructed to: 1. Immediately check your internal inventory to locate the product listed on the attached business reply form and remove them from their point of use. 2. Return the enclosed business reply form by email to RSRecall@Stryker.com to confirm receipt of this notification. 3. Upon receipt of the completed business reply form, Stryker will contact you to arrange for the return and replacement of your product(s). 4. Maintain awareness of this communication internally until all required actions have been completed within your facility. Please keep this communication close to the affected items until replacement. 5. Inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients appropriately. For questions or concerns - contact Customer Service +1 800 787 9537, option 2, from 8:00 AM to 7:00 PM (Eastern Time), Monday Friday or by email at medtechsup@stryker.com.
- Root cause
- Device Design
- Status
- Open, Classified
- Product code
- MKJ
- Quantity affected
- 622
- Distribution
- U.S. Nationwide distribution including in the states of CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, MA, MD, MI, MO, MS, NC, NH, NY, OH, PA, SC, TN, TX, VA, WA, WI, and WV.
- Date initiated
- 2025-01-21
- Date posted
- 2025-02-19
- Location
- Redmond, WA