Leksell Vantage Stereotactic System. Instrument Carrier, Part of Leksell Vantage Arc (1053958). Neurological stereotaxic

FDA device recall Z-1826-2020 · posted 2020-04-29 · Terminated

Recall details

Recalling firm
Elekta Inc
Reason for recall
A faulty locking piece of the Instrument Carrier that does not fulfil the requirement of locking force to the Arc has been found during an internal check of a Leksell Vantage Stereotactic Arc System.
Action taken
Elekta Inc., notified customers on 04/01/2020 via Urgent Important Field Safety Notification letter. The letter identified the affected products and customers were instructed to complete and return the Important Field Safety Notice Acknowledgement form to their local Elekta representative as soon as possible but no later than 30 days from receipt. Elekta will replace all defective Instrument Carriers. Replacement units will be sent to affected facilities as soon as available. The firm sent an additional letter on 05/06/2020 with the following information: All Instrument Carriers included in this IFSN will be replaced by Elekta. While you await your replacement, Elekta has formulated an additional mandatory in-field test that must be carried out on every defective device. This test must be carried out by a trained technician/engineer or equivalent at your hospital or by Elekta, and the defective device must pass the full test for the continued use of the product. Thus, if your defective Instrument Carrier should fail to pass this test, it must be immediately removed from use.
Root cause
Process design
Status
Terminated
Product code
HAW
Quantity affected
25 units.
Distribution
US nationwide distributions in the states of MA, NH, NC, UT, OH, FL, MN, NY, LA and MT.
Date initiated
2020-03-30
Date posted
2020-04-29
Date terminated
2024-04-10
Location
Saint Charles, MO

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
Leksell Vantage Stereotactic System (K190887)Elekta Instrument AB2019-08-01
Leksell Vantage Stereotactic System (K171123)Elekta Instrument AB2017-09-15