Leksell GammaPlan 8.0 image fusion Leksell GammaPlan is designed for use with the Leksell Gamma Knife manufactured by El
FDA device recall Z-1604-2010 · posted 2010-05-14 · Terminated
Recall details
- Recalling firm
- Elekta, Inc.
- Reason for recall
- The precision of the calculation used to create the fused study in LGP 8.0 is too low and should not be used until the system is upgraded to LGP 8.2.
- Action taken
- Phase 1 - Important Notice and Technical Note #200 058 - "Accuracy problems in Leksell GammaPlan 8.0 image fusion", dated December 10.2007 and Field Change Order #200 058, "Technical Note: Accuracy problems in Leksell Gamma Plan 8.0 image fusion", dated January 7, 2008 were sent to affected customers and all field service engineers as notification of the problem. The notice describes the product, problem and actions to be taken by customers. The firm recommended that the customers not use the fusion function until their Leksell GammaPlan(R) is upgraded to version 8.2. For further question, do not hesitate to contact your Elekta support respresentative at
- Root cause
- Other
- Status
- Terminated
- Product code
- IWB
- Quantity affected
- 66 units
- Distribution
- Nationwide distribution: AL, AR, CA, CO, CT, FL, GA, IL MD, MA, MI, MN, MS, MO, NJ, NM, NY, NC, OH, OK, OR, PA, SC, TN, TX, UT, VA, and WA.
- Date initiated
- 2008-01-07
- Date posted
- 2010-05-14
- Date terminated
- 2010-12-28
- Location
- Norcross, GA