Leksell Gamma Knife. Leksell Gamma Knife is a teletherapy device indicated for use in the stereotactic irradiation of in

FDA device recall Z-1236-2010 · posted 2010-04-13 · Terminated

Recall details

Recalling firm
Elekta, Inc.
Reason for recall
After updating the LGK actuator in spare part 810361, the old sleigh became obsolete due to causing insufficient locking of the helmet in combination with the new actuator.
Action taken
Field Change Order 0000203 "Replacement of obsolete actuator sleigh on Helmet Changer", was sent to all customers that were affected. Direct questions to Elekta, Inc. by calling 1-770-670-2548.
Root cause
Other
Status
Terminated
Product code
IWB
Quantity affected
5 units
Distribution
United States (CA, TN, TX and WA) and Taiwan.
Date initiated
2007-10-01
Date posted
2010-04-13
Date terminated
2010-12-28
Location
Norcross, GA