Leica M822 Surgical Microscope Device Controller Software (MDC version 3.6, pack F, affecting the control unit component

FDA device recall Z-1042-2012 · posted 2012-02-16 · Terminated

Recall details

Recalling firm
Leica Microsystems, Inc.
Reason for recall
Leica Microsystems received complaints stating that the zoom function of the Leica M822 surgical microscope locked up during use.
Action taken
The firm, Leica Microsystems, Inc., sent a "Medical Device Correction" notice dated December 21, 2011 to all affected customers by e-mail. The product, problem, and actions to be taken by the customers were addressed. Customers were instructed- DO NOT USE use the affected product until the software could be upgraded and complete and return the Acknowledgement Form via fax to: 1-847-607-3147 or Scan and send to elizabeth.culotta@leica-microsystems.com. Leica Microsystems representatives will contact the customers to make necessary arrangements to complete the software upgrade. If you have any questions, contact Director, Global Product Quality Management at 1-847-317-7209.
Root cause
Device Design
Status
Terminated
Product code
ETP
Quantity affected
29 units (10 units-USA)
Distribution
Worldwide distribution: USA (nationwide) including states of: California, Colorado, Illinois, New York, Oregon, South Carolina; and countries of: Australia, Japan, Korea, Pakistan, Romania, South Afri
Date initiated
2011-12-21
Date posted
2012-02-16
Date terminated
2012-02-16
Location
Buffalo Grove, IL