Leica M525 F50 Surgical Microscope; Article No. 10338422; Leica Microsystems (Schweiz) Ag, 201 Max Schmidheiny Strasse,
FDA device recall Z-2053-2012 · posted 2012-07-20 · Terminated
Recall details
- Recalling firm
- Leica Microsystems, Inc.
- Reason for recall
- Leica Microsystems evaluated data showing that a malfunction resulting in an illumination reduction can be observed when Leica M525 F50 surgical microscope is used in combination with a remote control connected to a remote video adapter or motorized dual video adapter.
- Action taken
- Leica Microsystems sent an Urgent Medical Device letter dated June 4, 2012 to the affected customers. The letter described the product, problem and actions to be taken. The customers were informed that the Leica surgical microscope M525 F50 required a software upgrade. Customers were instructed to NOT USE a remote video adapter or motorized video adapter until the required software upgrade has been installed by their Leica Microsystems representative. The customers were requested to complete the attached Acknowledgement Form, acknowledging receipt and understanding of the letter, listing the serial numbers of the units at their facilities and to fax it back to Leica Microsystems at (847) 236-3747.
- Root cause
- Software design
- Status
- Terminated
- Product code
- FSO
- Quantity affected
- 39 Units - USA , 4 Units - Canada
- Distribution
- Worldwide Distribution - US (nationwide) in the states of: California, Colorado, Florida, Georgia, Illinois, Indiana, Massachusetts, Michigan, Minnesota, New York, Oklahoma, Oregon, Pennsylvania, Texa
- Date initiated
- 2012-06-04
- Date posted
- 2012-07-20
- Date terminated
- 2012-10-29
- Location
- Buffalo Grove, IL