Atlas FDA

Leica Biosystems Cryostat Model Number CM1900, Material Numbers: a) 14045227382, b) 14901900100, c) 14901900700, d) 1491

FDA device recall Z-0088-2025 · posted 2024-10-15 · Open, Classified

Recall details

Recalling firm
LEICA BIOSYSTEMS NUSSLOCH GMBH
Reason for recall
As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
Action taken
Leica Biosystems issued an Urgent Field Action Notice to its consignees on 09/05/2024 via mail. The notice explained the problem with the device and requested the placement of a warning sticker regarding the use of flammable freezing sprays on the device and the addition of a supplemental sheet to the current IFU. For questions, please contact: Heather Jensen, Leica Biosystems, 5205 US Highway 12, Richmond, IL 60071 USA, Tel.: +1-815-298-8371, heather.jensen@leicabiosystems.com
Root cause
Labeling design
Status
Open, Classified
Product code
IDP
Quantity affected
25 units
Distribution
US Nationwide distribution.
Date initiated
2024-09-04
Date posted
2024-10-15
Location
Nussloch, Germany