Atlas FDA

LATITUDE NXT Remote Patient Management System, LATITUDE NXT System Server Software Model 6460.

FDA device recall Z-1216-2023 · posted 2023-03-03 · Open, Classified

Recall details

Recalling firm
Boston Scientific Corporation
Reason for recall
Under specific circumstances, the U.S. product registration system did not send up enablement requests to U.S. LATITUDE Customer Support during an 8-month interval in 2022. For these patients, the HeartLogic Index and any associated Yellow Alerts are currently not visible in LATITUDE as may have been intended by the healthcare provider..
Action taken
The recalling firm issued letters to LATITUDE Healthcare Professionals via third-party Federal Express with signature at the account required on 2/2/2023. The letter informed the HCP that records indicate they may have U.S. patients who do not have HeartLogic enabled as the HCP may have intended in the Model 6460 U.S. LATITUDE NXT Patient Management System. The patients included within the enclosed list are scheduled to have HeartLogic enabled in LATITUDE ON 2/14/2023. The letter provides the clinical impact and recommendations. The recommendations include: (1) Distribute this information to appropriate HCPs within your center who manage HeartLogic alerts for the patients in the enclosed list; and (2) If you do NOT want HeartLogic enabled in LATITUDE for patients in the enclosed list, contact Boston Scientific LATITUDE Customer Support via latitude@bsci.com before 2/14/2023 and identify which patients should NOT be enabled. On 2/14/2023, BSC will correct the data on the Model 6460 LATITUDE Server for affected patients, unless a request not to enable is received by the firm prior to this date. Update: on 02/14/2023, Boston Scientific sent an additional letter, due to the original letter being lost in the mail. The revised letter advised customers to email LATITUDE Custoemr Support to have HeartLogic disabled.
Root cause
Software Manufacturing/Software Deployment
Status
Open, Classified
Product code
LWS
Quantity affected
1,531 patient data records
Distribution
US Nationwide distribution to the healthcare facilities and healthcare providers are located in CA, CO, DE, FL, GA, ID, IL, IN, IA, KY, ME, MD, MA, NV, NH, NJ, NY, NC, OH, PA, RI, SC, TN TX, VA, WA, W
Date initiated
2023-02-02
Date posted
2023-03-03
Location
Saint Paul, MN