LACTOSORB SYSTEM 2.0MM ADJUSTABLE SELF-DRILLING TAP, REF: 915-2075, NON-STERILE PRODUCT, BIOMET MICROFIXATION,K INC. 152
FDA device recall Z-0489-2010 · posted 2010-01-04 · Terminated
Recall details
- Recalling firm
- Biomet Microfixation, Inc.
- Reason for recall
- Biomet Microfixation is recalling LactoSorb Self Drilling Adjustable Tap Model Numbers 915-2075 Lot # 064350 and 915-2185 Lot # 885580. The two lots identified have a new style stop and an old style tap; they are not compatible, as the depth etchings do not line up correctly. Depending on the tap (1.5mm VS 2.0mm) you may drill too shallow or too deep.
- Action taken
- Biomet Microfixation notified consignees of the recalled product beginning September 2009 via telephone and e-mail correspondence. Users were asked to return the affected product to the firm. For further information, contact Biomet Microfixation at 1-904-741-9468.
- Root cause
- Component change control
- Status
- Terminated
- Product code
- HWC
- Quantity affected
- 15 pieces
- Distribution
- USA - FL, MO, and KS; Foreign: Brazil and China.
- Date initiated
- 2009-09-15
- Date posted
- 2010-01-04
- Date terminated
- 2010-01-23
- Location
- Jacksonville, FL
Monitor recalls programmatically