Kinamed SuperCable Grip and Plate System Driver, 3.5mm Hex, Catalog Number: 35-860-2060 The driver is intended to be use

FDA device recall Z-2588-2011 · posted 2011-06-17 · Terminated

Recall details

Recalling firm
Kinamed Inc
Reason for recall
The recall was initiated due to inadequate mechanical strength of these screwdrivers' hexagonal tip feature, causing it to strip during final tightening.
Action taken
The firm, Kinamed Inc., sent a recall letter dated August 18, 2010 to consignees/customers via FedEx. The letter described the product, problem and actions to be taken. The customers were instructed to Immediately examine their stock and return any Kinamed SuperCable Screwdriver, 3.5mm Hex bearing these lot numbers, along with the completed attached Effectiveness Check Form and shipment under RMA #74567 to the attention of: "Director of Quality Assurance" Kinamed, Inc. 820 Flynn Road Carnarillo, CA 93012-8701 Note: The Kinamed sales agent shipped and had the product replaced for the customers. If you have any questions, please contact (303)-681-2208, (805) 384-2748 or (800) 827-5775
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
KTT
Quantity affected
8 units
Distribution
Worldwide distribution: USA (nationwide) state of: CO and country of: Columbia.
Date initiated
2010-08-18
Date posted
2011-06-17
Date terminated
2011-10-25
Location
Camarillo, CA