Kenex Radiation Shield and surgical lamps installed with GE Healthcare Interventional, Surgery, CT and X-Ray Imaging sys
FDA device recall Z-1706-2014 · posted 2014-06-03 · Terminated
Recall details
- Recalling firm
- GE Healthcare, LLC
- Reason for recall
- GE Healthcare has recently become aware of a potential safety issue due to the improper installation of a Kenex ceiling suspended radiation shield and surgical lamp. This issue is not related to the design of the Kenex product. The affected products are limited to installations completed or contracted by GE Healthcare. There has been a reported incident of a radiation shield falling from the ceil
- Action taken
- GE Healthcare sent an "Urgent Medical Device Correction" letter # 12209 dated May 2, 2014. The letter was addressed to Director of Biomedical Engineering, Director of Radiology and Chief of Cardiology. The letter described the Safety Issue, Safety Instructions, Affected product Details, Product Correction and Contact Information. For questions they should call 1-800-345-2700. For other countries, they should contact their local GE Healthcare Service Representative. .
- Root cause
- Equipment maintenance
- Status
- Terminated
- Product code
- KPY
- Quantity affected
- 3221 (108 US, 3113 OUS).
- Distribution
- Worldwide Distribution - US Nationwide, ALBANIA, ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BOLIVIA, BOSNIA & HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHILE,
- Date initiated
- 2014-05-02
- Date posted
- 2014-06-03
- Date terminated
- 2015-02-25
- Location
- Waukesha, WI