Atlas FDA

Juggerknot Mini Soft Anchors, Model/Catalog Number: 912080

FDA device recall Z-2619-2025 · posted 2025-09-23 · Open, Classified

Recall details

Recalling firm
Biomet, Inc.
Reason for recall
The outer carton box label is for the JuggerKnot 1.0mm Mini products. However, the products within the carton are JuggerKnot 1.4mm Short devices.
Action taken
An URGENT MEDICAL DEVICE RECALL notification letter dated 9/9/25 was sent to customers. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the recall notice and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Distributor Responsibilities: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form for each return and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form must be returned even if you do not have affected products available to return in your territory. b. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. c. M
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
MBI
Quantity affected
40
Distribution
US Nationwide distribution.
Date initiated
2025-09-09
Date posted
2025-09-23
Location
Warsaw, IN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
BIOMET SPORTS MEDICINE JUGGERKNOT (TM) MINI SOFT ANCHOR (K110879)Biomet Sports Medicine2011-08-19