JResultNet drivers using ASTM coding language. Calculator/data processing module, for clinical use - Stand-alone softwar

FDA device recall Z-1680-2017 · posted 2017-02-16 · Terminated

Recall details

Recalling firm
Data Innovations, LLC
Reason for recall
Software issue resulting in the results of one patient potentially being associated with another.
Action taken
Data Innovations, LLC contacted consignees by phone, email, and post. Recall notifications were sent out on November 3, 2014. Consignees were provided instructions on how to reconfigure the drive to work around the issue. If consignees were unable to perform the work around, Customer Support is able to service the units. In order to request this service, consignees were to call 239-931-6004, 802-658-1955, or 800-332-0499. Attached to the recall notification was a response form that is to be faxed to 802-658-2782. Consignees with and without affected drives are to complete this response form. For questions regarding this recall call 239-931-6004.
Root cause
Software design
Status
Terminated
Product code
JQP
Quantity affected
657 units
Distribution
Nationwide Distribution to AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, W
Date initiated
2014-10-23
Date posted
2017-02-16
Date terminated
2019-08-21
Location
South Burlington, VT