Jadak Barcode Scanner, Part No. 350587-32, 350587-41, 351377-32. Product Usage: An accessory for the Pyxis Anesthesia ES
FDA device recall Z-0814-2016 · posted 2016-02-16 · Terminated
Recall details
- Recalling firm
- CareFusion 303, Inc.
- Reason for recall
- Customers reported issues when scanning medications with the scanners. When a user scans a medication using the affected barcode scanner, information on a different medication could be communicated to the Pyxis product.
- Action taken
- BD sent an Urgent Medical Device Recall Notification letter dated January 15, 2015 to all affected customers The letter identified the affected product, problem and actions to be taken. Customers are instructed to complete the enclosed Customer Response Card by selecting the desired deployment option and return the card to CareFusion. Customers with questions are instructed to contact the CareFusion Support Center at (888) 562-6018. Customers with adverse event reports are instructed to contact CareFusion Customer Advocacy at (800) 836-7842. Customers with technical questions are instructed to contact CareFusion Technical Support at (800) 727-6102.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- BRY
- Quantity affected
- 9,345 units
- Distribution
- Worldwide:Distribution - US Nationwide and the countries of Saudi Arabia, Bahrain, Australia, United Arab Emirates, Qatar, Mexico, Guam and the Bahamas.
- Date initiated
- 2016-01-15
- Date posted
- 2016-02-16
- Date terminated
- 2017-05-08
- Location
- San Diego, CA