FDA device recall Z-0154-2026 · posted 2025-10-23 · Open, Classified

Recall details

Recalling firm
ICU Medical, Inc.
Reason for recall
IV Gravity burette administration set burette component is missing an internal shutoff valve intended to stop fluid flow, which if used may result in delay in therapy during setup, over delivery/unrestricted flow, or air may be infused into the body.
Action taken
ICU notified consignees via email with an "URGENT MEDICAL DEVICE CORRECTION" on about 10/02/2025. Consignees were instructed to: 1) Check all inventory locations within your institution for the affected products and discontinue use. Destroy all affected products following your institution s process for destruction. If destroying is not immediately possible at your facility, then the product should be quarantined until disposal. 2) Share this notification with all potential users of the device, to ensure they are aware of this notification. If the devices are used at another location, please ensure this communication is delivered there. 3) Complete and return the attached Customer Response Form to marketaction@mailac.custhelp.com within 10 days of receipt to acknowledge your understanding of this notification. 4) If you have distributed affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to marketaction@mailac.custhelp.com. ICU Medical will provide replacement products to affected customers upon receipt of a complete Customer Response Form to certify product destruction. If you have any questions, contact Customer Service via email: customerservice@icumed.com or call 1-(800)-258-5361 Additional information or technical assistance.
Root cause
Process control
Status
Open, Classified
Product code
FMG
Quantity affected
23203
Distribution
Worldwide distribution: US (nationwide) to states of: TX, SC and OUS (foreign) to countries of: Canada, Australia, and New Zealand.
Date initiated
2025-10-02
Date posted
2025-10-23
Location
San Clemente, CA