Isopure Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO) with a Finish Thompson Centrifugal Pump Model #11

FDA device recall Z-0822-2020 · posted 2020-01-19 · Terminated

Recall details

Recalling firm
Isopure Corp
Reason for recall
There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.
Action taken
The only affected customer was contacted via telephone on 4/19/19. An URGENT PRODUCT FIELD CORRECTION NOTICE dated 5/22/19 was distributed to affected clinics.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
FIN
Quantity affected
1
Distribution
The products were distributed to the following US states: AL, FL, GA, MA, MD, MN, NC, NM, NY, OH, OK, TX, and VA.
Date initiated
2019-04-19
Date posted
2020-01-19
Date terminated
2020-05-29
Location
Louisville, KY