ISOLOC Software Version 6.0, Part Number NT-NW-425-520. Localization software for radiotherapy treatment system.

FDA device recall Z-1154-2007 · posted 2007-08-08 · Terminated

Recall details

Recalling firm
Northwest Medical Physics Equipment Inc
Reason for recall
When ISOLOC version 6.0 software was used in the Anatomic Landmark mode, the couch moves display in the Couch Translator window was incorrect.
Action taken
Software to fix the described problem was written and customers contacted by e-mail on 12/30/2004 that a software patch, titled 12302004 was available to their customers on the firm's internet website.
Root cause
Software design
Status
Terminated
Product code
IYE
Quantity affected
79
Distribution
Worldwide, including USA, Korea, Spain, and UK.
Date initiated
2004-12-30
Date posted
2007-08-08
Date terminated
2011-04-19