ISOLOC Software Version 6.0, Localization software for determining treatment machine position (to position patients) for

FDA device recall Z-1153-2007 · posted 2007-08-08 · Terminated

Recall details

Recalling firm
Northwest Medical Physics Equipment Inc
Reason for recall
When ISOLOC version 6.0 was used in the centimeter mode, the display in the couch moves window had a rounding error relative to the Portal Image Marker window of up to 0.5 mm.
Action taken
Software to fix the described problem was written and customers contacted by e-mail on 04/11/2005, that a software patch, titled 04112005, was available to their customers on the firm's internet website.
Root cause
Software design
Status
Terminated
Product code
IYE
Quantity affected
84
Distribution
Worldwide, including USA, Korea, Spain, and UK.
Date initiated
2005-04-11
Date posted
2007-08-08
Date terminated
2011-04-19
Location
Everett, WA