ISOLOC Software Version 4.2, Part Number NT-NW-425-520. Image guided radiation treatment software.

FDA device recall Z-1155-2007 · posted 2007-08-09 · Terminated

Recall details

Recalling firm
Northwest Medical Physics Equipment Inc
Reason for recall
When ISOLOC Version 4.2 unit is set to centimeter and the user clicks the Reenter button in the digitizer window, the coordinates of the markers and target in the anatomic imaging system erroneously displays values ten times greater than the correct values.
Action taken
Software to fix the described problem was written and customers contacted by e-mail on 11/30/2004, that a software patch, titled 11302004, was available to their customers on the firm's internet website.
Root cause
Software design
Status
Terminated
Product code
IYE
Quantity affected
83
Distribution
Worldwide, including USA, Korea, Spain, and UK.
Date initiated
2004-11-30
Date posted
2007-08-09
Date terminated
2011-04-19
Location
Everett, WA