Atlas FDA

IsoLoc Prostate Immobilization Treatment Balloon Device Kit, Catalog/UPN: ISO-100

FDA device recall Z-2267-2024 · posted 2024-07-03 · Open, Classified

Recall details

Recalling firm
Angiodynamics, Inc.
Reason for recall
A single batch of ISO-100 IsoLoc Gas-Release Endorectal Balloons is mislabeled. The affected ISO-100 balloon catheter has a nominal fill volume of 100cc/ml but the inner pouch identifies the product as an ISO-60 (fill volume of 60cc/ml).
Action taken
On May 21, 2024 URGENT VOLUNTARY MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: IMMEDIATELY: 1. Stop using the product subject to recall. 2. Remove any affected (recalled) product from your inventory (whether in labs, Central Supply, Shipping and Receiving or ANY other location). 3. Segregate this product in a secure location for return to AngioDynamics, Inc. 4. Forward a copy of this recall notification to all sites to which you have distributed affected product.
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
PCT
Quantity affected
1,500 units (75 kits with 20 units per kit)
Distribution
US Nationwide distribution in the states of AZ, FL, IL, MO, MS, NJ, NM, NY, PA, TX, WI.
Date initiated
2024-05-21
Date posted
2024-07-03
Location
Queensbury, NY

Monitor recalls programmatically