ISOLOC 4.2 Software Part Number NT-NW-425-520. Image guided radiation treatment software.

FDA device recall Z-1147-2007 · posted 2007-08-08 · Terminated

Recall details

Recalling firm
Northwest Medical Physics Equipment Inc
Reason for recall
When the ISOLOC (Version 4.2) profile distance unit is set to "centimeter" and the user clicked the "back" button in the digitizer window, the marker and target coordinates in the localization window erroneously display values ten times greater than the correct values.
Action taken
Software to fix the described problems was written and customers contacted by e-mail on 05/31/2005, that a software patch, titled 05312005 was available on the firm's internet website.
Root cause
Software design
Status
Terminated
Product code
IYE
Quantity affected
85
Distribution
Worldwide, including USA, Korea, Spain, and UK.
Date initiated
2005-05-31
Date posted
2007-08-08
Date terminated
2011-04-19
Location
Everett, WA