Atlas FDA

Isoflurane E_Vap.(Safety Filling adapter): used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic

FDA device recall Z-2951-2024 · posted 2024-09-04 · Open, Classified

Recall details

Recalling firm
Mindray DS USA, Inc. dba Mindray North America
Reason for recall
V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention
Action taken
Mindray issued an "URGENT Medical Device Recall" letter on July 22, 2024. Letter states reason for recall, health risk and actions to be taken: The V90 electronic vaporizer can continue to be used normally. If "Vaporizer Output Abnormal" alarm or other alarms associated with abnormal vaporizer output or high anesthetic agent alarms occur during use, take appropriate intervention per the A9 Operators manual and contact Mindray Technical Support for assistance. The intervention could be switching to another V90 electronic vaporizer, or the user can decide to switch to intravenous anesthesia. For vaporizer replacement, refer to A9 Operators manual part number H-046-017199-00, section 3.4. Vaporizer . The manual is available on the Mindray website. Mindray s Technical Support team can assist and may be reached at (877) 913-9663 (Option 1) Monday through Friday, 8:30 a.m. 5:30 p.m. ET. Please ensure that applicable clinical staff are aware of the content of this letter. Mindray will replace all affected V90 electronic vaporizers located at your facility. The V90 electronic vaporizers will be shipped directly to your facility, and a Mindray Care Team representative will contact you to schedule the installation and collect the affected V90 electronic vaporizers Contact: Director, Quality Operations and Regulatory Affairs US Agent for Shenzhen Mindray Biomedical Electronics Corp., Ltd Mindray DS USA, Inc. 201.995.8407
Root cause
Nonconforming Material/Component
Status
Open, Classified
Product code
BSZ
Quantity affected
24 units
Distribution
US Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: Canada
Date initiated
2024-07-22
Date posted
2024-09-04
Location
Mahwah, NJ

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
A8, A9 Anesthesia System (K201957)Shenzhen Mindray Bio-Medical Electronics Co., Ltd.2021-03-26