iSite PACS image management system; Picture Archiving and Communications System; Radiology software package, Philips Hea

FDA device recall Z-1141-2008 · posted 2008-05-08 · Terminated

Recall details

Recalling firm
Philips Healthcare Informatics, Inc.
Reason for recall
Incorrect display of one of the patient's images. This malfunction could cause an image to be incorrectly displayed, for example flipped left to right and anterior to posterior. The original image is still presented in the correct orientation next to the incorrect image when displayed on the monitor.
Action taken
Philips Healthcare Informatics mailed customers notification: Urgent Product Correction regarding FCO#2954704-01/04/08-001-C. Firm released a corrected version of software (iSite PACS v.3.5.69) on January 21, 2008. Firm is scheduling customers to upgrade to the corrected version.
Root cause
Software design
Status
Terminated
Product code
LLZ
Quantity affected
197 units
Distribution
Worldwide Distribution: USA, United Arab Emirates, Germany, the Netherlands, Denmark, Australia, France, Italy, Belgium, Ireland, Italy, and Canada.
Date initiated
2008-01-04
Date posted
2008-05-08
Date terminated
2009-10-07
Location
Foster City, CA