ISE REF Valve, Part No. MU3226 for the AU400, AU640, AU2700, and AU5400 Chemistry Analyzer. Automated chemistry analyzer

FDA device recall Z-0152-2014 · posted 2013-11-05 · Terminated

Recall details

Recalling firm
Beckman Coulter Inc.
Reason for recall
The recall was initiated because Beckman Coulter has confirmed that some AU Clinical Chemistry Analyzer ISE units can experience premature ISE REF (Reference) valve failure. Premature valve failure can be detected by the presence of air bubbles in the ISE REF solution tubing on the ISE unit. If air bubbles are present, accuracy of the ISE test results may be affected.
Action taken
A customer notification letter dated 9/9/13 was sent to customers who received the ISE REF Valve, Part Number MU3226 for the AU400, AU640, AU2700, AU5400 and the ISE REF Valve, Part Number MU7638 for the AU480, AU680, AU5800 Chemistry Analyzers. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with questions are instructed to contact Beckman Coulter Customer Support Center via http://www.beckmancoulter.com/customersupport/support, via phone at (800) 854-3633 in the US and Canada. Outside of the US and Canada, customers are instructed to contact their local Beckman Coulter Representative.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
JJE
Quantity affected
4584 units total (1173 units in US)
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Canada, Vietnam, Zambia, and Zimbabwe.
Date initiated
2013-09-09
Date posted
2013-11-05
Date terminated
2014-09-30
Location
Brea, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
OLYMPUS AU5400 CLINICAL CHEMISTRY ANALYZER (K011720)Olympus America, Inc.2001-06-25
MODIFICATION TO OLYMPUS AU2700 CLINICAL CHEMISTRY ANALYZER (K003721)Olympus America, Inc.2000-12-18
OLYMPUS AU400 CLINICAL CHEMISTRY ANALYZER (K981743)Olympus America, Inc.1998-06-09
OLYMPUS AU600 CLINICAL CHEMISTRY ANALYZER (K961274)Olympus America, Inc.1996-07-05