IR remote control JUX, Material no.1307314, Serial number: to (incl.) 102338310 Patient positioning during surgery, rang

FDA device recall Z-0927-2015 · posted 2015-01-06 · Terminated

Recall details

Recalling firm
Trumpf Medical Systems, Inc.
Reason for recall
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes
Action taken
Trumpf sent an Urgent Medical Device Correction letter on December 3, 2014, to all affected customers. Customers were instructed to complete and return the Confirmation of Receipt form with serial numbers of the affected table(s) and the serial number(s) for any remote controls they have. Replacement remotes will be sent. Customers were instructed to dispose of their recalled remotes. Customers with questions were instructed to call 888-474-9359. For questions regarding this recall call 843-534-0606.
Root cause
Component change control
Status
Terminated
Product code
JEA
Quantity affected
3303 total all devices
Distribution
Worldwide Distribution - USA including AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MK, MS, NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI and Internationally to Canada.
Date initiated
2014-12-03
Date posted
2015-01-06
Date terminated
2016-08-23
Location
Charleston, SC