IR remote control JUPITER U, Material no. 4500120, Serial number: to (incl.) 102306940. Patient positioning during surge
FDA device recall Z-0921-2015 · posted 2015-01-06 · Terminated
Recall details
- Recalling firm
- Trumpf Medical Systems, Inc.
- Reason for recall
- Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes
- Action taken
- Trumpf sent an Urgent Medical Device Correction letter on December 3, 2014, to all affected customers. Customers were instructed to complete and return the Confirmation of Receipt form with serial numbers of the affected table(s) and the serial number(s) for any remote controls they have. Replacement remotes will be sent. Customers were instructed to dispose of their recalled remotes. Customers with questions were instructed to call 888-474-9359. For questions regarding this recall call 843-534-0606.
- Root cause
- Component change control
- Status
- Terminated
- Product code
- JEA
- Quantity affected
- 3303 total all devices
- Distribution
- Worldwide Distribution - USA including AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MK, MS, NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI and Internationally to Canada.
- Date initiated
- 2014-12-03
- Date posted
- 2015-01-06
- Date terminated
- 2016-08-23
- Location
- Charleston, SC