Atlas FDA

iQ200 Series Urine Microscopy Analyzer: a) iQ 200 Select, Part Numbers C10684 and 700-3345; b) iQ 200 Elite, Part Number

FDA device recall Z-1596-2023 · posted 2023-05-17 · Open, Classified

Recall details

Recalling firm
Beckman Coulter, Inc.
Reason for recall
Beckman Coulter has become aware of an intermittent issue in which the optional flag Previous Sample Had Sperm was enabled but not displayed so that a carryover event could have been investigated.
Action taken
Beckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 04/05/2023 by mail. The notice explained the issue and the hazard and requested the following actions be taken: "Examine your specimen settings to determine if the optional 'Sperm Present' and 'Previous Sample Had Sperm' flags are enabled. If disabled, no further action is needed by your laboratory. If enabled, follow the actions below If the presence of sperm is identified: Review the previous specimen for the presence of sperm. Follow recommendations under Previous Sample Had Sperm in the iQ200 Series Instructions For Use (IFU) 300-4320CE and 300-4321EE and DxU 850m and DxU 840m Iris IFU (C49320AB) in Chapter 6, Data Review, Flags, Previous Sample Had Sperm, Recommendations. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter."
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
KQO
Quantity affected
1068 units
Distribution
Worldwide Distribution
Date initiated
2023-04-05
Date posted
2023-05-17
Location
Miami, FL

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
iQ200 System, iChemVELOCITY Automated Urine Chemistry System (K210127)Beckman Coulter, Inc.2021-07-02
IQ 20 SERIES ANALYZER, IQ 200 SERIES SYSTEM, NAD IQ LAMINA CRADLE (K093861)Iris International, Inc.2010-02-05
IQ 200 SYSTEM (K022774)Intl. Remote Imaging Systems2002-10-21