Intuitive Surgical EndoWrist, 8 mm, PK Dissecting Forceps for use with the da Vinci and da Vinci S Surgical systems, Par

FDA device recall Z-0258-2008 · posted 2008-01-24 · Terminated

Recall details

Recalling firm
Intuitive Surgical, Inc.
Reason for recall
Mislabeling-electrical isolation requirements: devices were incorrectly labeled with a CF symbol not their proper BF Symbol on the instrument housing. (All instruments meeting the CF isolation requirements are suitable for direct cardiac application, while instruments meeting BF isolation requirements are not suitable for direct cardiac application)
Action taken
On January 9, 2008, via an Urgent Device Recall Letter the recalling firm notifed users of the defect (mislabeling with heart symbol) and to Immediately Discontinue any direct cardiac applications. The recalling firm will arrange with each consignee a time to relabel their devices. Customers on April 17, 2007 had earlier been warned of the device defect via letter but no correction was indicated.
Root cause
Device Design
Status
Terminated
Product code
NAY
Quantity affected
1,136 units
Distribution
Worldwide Distribution --- USA including states of MN, PA, AK, NY, CA, MD, IN, AZ, CO, AR, TX, TN, FL, VA, MA, NE, NC, IL, OH, ME, WI, MD, NJ, MN, MI, VA, IA, OR, MO, NC, MS, AL, WA, GA, KY, LA, AK, A
Date initiated
2007-04-13
Date posted
2008-01-24
Date terminated
2008-08-01
Location
Sunnyvale, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
INTUITIVE SURGICAL ENDOWRIST PK DISSECTING FORCEPS, MODELS 400214 & 420214 (K061260)Intuitive Surgical, Inc.2006-05-18