International Technidyne Corp. (ITC) AVOXimeter 4000 Co-oximeter, Catalog Numbers: AVOX4000-110 and AVOX4000-220

FDA device recall Z-1239-2008 · posted 2008-08-06 · Terminated

Recall details

Recalling firm
International Technidyne Corp.
Reason for recall
Incorrect reading: ITC discovered that instruments built or repaired since approximately April 2007 were inadvertently left with the diagnostic mode enabled when shipped to customers.
Action taken
ITC sent Urgent - Medical Device Field Recall - ITC No. 08-001 letters dated 2/08/2008 to hospitals and/or medical facilities identified as having affected instruments. The letter instructed users not to use the instrument if the diagnostic mode is enabled. ITC will reset affected devices and return them to the customers.
Root cause
Process design
Status
Terminated
Product code
JKS
Quantity affected
61 US, 2 International
Distribution
Worldwide Distribution: USA, United Arab Emirates, Malaysia, Saudi Arabia, India, Germany, United Kingdom, Greece, Romania, Singapore, ROC, Pakistan, Netherlands, Korea, France, Belgium, Switzerland,
Date initiated
2008-02-13
Date posted
2008-08-06
Date terminated
2009-07-24
Location
Edison, NJ