Atlas FDA

Intera 1.5T Master/Nova, Model Number: 781106;

FDA device recall Z-1776-2025 · posted 2025-05-15 · Open, Classified

Recall details

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Reason for recall
The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
Action taken
Philips originally notified consignees on 03/24/2025 with Updated URGENT Medical Device Correction letter. It was noted that it contained incorrect information and on 04/09/2025 Philips sent an updated Updated URGENT Medical Device Correction letter. Consignees were instructed to take the following actions prior to a patient scan: 1. Inspect the QBC seal for separation between the cone cover and QBC cover. 2. If QBC seal is found detached or loose, Stop-use immediately. 3. Contact your local Philips service representative. If QBC seal becomes loose during a patient scan: 1. Immediately stop scanning and carefully remove patient from the system. 2. Contact your local Philips service representative. Consignees were also instructed to circulate this URGENT Medical Device Correction letter to all users of this device so that they are aware of the issue, retain this letter with your system(s) until a solution is installed on your system, ensure the letter is in a place likely to be seen/viewed, and complete and return the included response form to Philips promptly and no later than 30 days from receipt via email to: philips.recall@philips.com.
Root cause
Process control
Status
Open, Classified
Product code
LNH
Quantity affected
103 units (37 US, 66 OUS)
Distribution
US: Nationwide Distribution International: Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Botswana, Brazil, Bulgaria, Canada, Cayman
Date initiated
2025-03-24
Date posted
2025-05-15
Location
Best, Netherlands

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
Ingenia 3.0T, Ingenia 3.0T CX, Ingenia Elition and MR 7700 with Distributed Multi Nudclei (K213516)Philips Medical Systems Nederland B.V.2022-03-03
Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia Ambition MR Systems (K193215)Philips Medical Systems Nederland B.V.2020-04-10