Intellispace Portal software Product Usage: Intellispace Portal is a multimodality thin-client applications server that
FDA device recall Z-2000-2012 · posted 2012-07-12 · Terminated
Recall details
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Reason for recall
- In the Multi-Modality Tumor Tracking application, when changing the contours of the lesion using the editing mode "edit contour" tool, the measurements are not recalculated causing discrepancies with the lesion measurements.
- Action taken
- Philips Healthcare sent an Urgent-Medical Device Correction letter dated May 23, 2012, via certified mail to each consignee. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review the information contained in the letter with all members of their staff who need to be aware of the contents of the communication. The letter informed affected customers that a Philips Service Engineer will contact them for implementation of the software update on the affected systems free of charge. For any further information or support, customers can contact their local Philips representative or local Philips Healthcare office. For North America and Canada contact the Customer Care Solutions Center (1-800-722-9377, option 5: Enter Site ID or follow the prompts).
- Root cause
- Software design
- Status
- Terminated
- Product code
- JAK
- Quantity affected
- 44
- Distribution
- Worldwide Distribution - US (nationwide) including the states of: AR, LA, MI, NC, NY, OH, OK and PA and the countries of: Australia, Canada, Denmark, France, Germany, Israel, Netherlands, Spain and Sw
- Date initiated
- 2012-05-22
- Date posted
- 2012-07-12
- Date terminated
- 2013-03-28
- Location
- Cleveland, OH
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| BRILLIANCE VOLUME (K060937) | Philips Medical Systems (Cleveland), Inc. | 2006-06-05 |
| PHILIPS PLUS COMPUTED TOMOGRAPHY SYSTEM (K033326) | Philips Medical Systems, Inc. | 2003-10-29 |
| MX8000 IDT CT SYSTEM (K012009) | Marconi Medical Systems, Inc. | 2001-09-13 |