Atlas FDA

InteliFUSE, Drill Bit, 2.00 mm, J-Latch, QTY: 1, REF: DB.200.JL, non-sterile, 1515 Poydras st., Ste 1490, New Orleans, L

FDA device recall Z-1033-2009 · posted 2009-03-05 · Terminated

Recall details

Recalling firm
Intelifuse Inc
Reason for recall
Manufacturing specification for drill bit was incorrect which, when used in conjunction with the drill guide, could cause the production of metal shavings.
Action taken
The firm initiated their recall on July 31, 2008 via telephone and Urgent Recall Notice to their current distributor, Integra. The letter asked them to complete, sign, and return the enclosed Recall Acknowledgment and Return Form. The letter also stated that InteliFUSE would travel to their designated Distribution Center(s) to ensure testing and replacement of all defective units once the form has been received. The firm sent follow-up Urgent Recall Notice letters on August 14, 2008 directing the distributor to recall from the hospitals to whom they distributed the product and to two hospitals which were shipped the products directly. The letter asked that use of the affected products be immediately ceased and that the any stock on hand be quarantined. The Recall Acknowledgment and Return Form should be completed, signed, and returned and the products should be returned to InteliFUSE for immediate replacement.
Root cause
Process control
Status
Terminated
Product code
HTW
Quantity affected
2149 units
Distribution
Nationwide Distribution - NY and NJ.
Date initiated
2008-07-30
Date posted
2009-03-05
Date terminated
2011-06-22
Location
New Orleans, LA