Atlas FDA

INTELECT VET, Model No. 2756, Chattanooga Group, Hixson, TN 37343, Made in the USA

FDA device recall Z-1223-2008 · posted 2008-06-25 · Terminated

Recall details

Recalling firm
Chattanooga Corp Inc
Reason for recall
Device malfunction: causing electric shock and burn.
Action taken
Beginning on 12/17/2007, Chattanooga Corp. sent two recall notification letters, via UPS mail, to its consignees informing them of the situation. The letters stated that the recall was to the user level and requested that the user perform the software upgrade, which will eliminate the possibility of shock and burn.
Root cause
Device Design
Status
Terminated
Product code
MPH
Quantity affected
140 units
Distribution
Worldwide Distribution - including USA
Date initiated
2007-12-17
Date posted
2008-06-25
Date terminated
2010-02-19
Location
Hixson, TN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
VECTRA EPR SYSTEM AND INTELECT EPR SYSTEM (K031077)Chattanooga Group2003-12-11